industrial furnace

55 Northeastern Blvd,Nashua, NH 03062sales@centorr.com
Ph: 603-595-7233Toll free: 800-962-8631Fax: 603-595-9220

USA

Chinese

Japanese

German
 
About Us | Products | Contact Us | Press Releases | Site Map | International Sales Reps
Career Opportunities | Parts & Service | Used Furnaces | Quarterly Sale
HOME
Series LF

Top Loading Multi-Purpose Furnace, High Temperature, Vacuum or Controlled Atmosphere, Refractory Metal Heat Zone

Professional and Uncluttered Appearance
 
Heat Zone
Heating Element
Tungsten/Molybdenum Heat Shields


  • Maximum Temperature: 2,000° C (3632° F)
  • Primary Use: Laboratory
  • Heat method: Electric Resistance
  • Operation: Vacuum and Inert Atmosphere
  • Operating Mode: Batch
  • Chamber Wall: Cold (Water Jacket)
  • Maximum Vacuum: 10-3 or 10-6 Torr (dependent on model selected)
  • Maximum Positive Pressure Inert Gas: 2 PSIG (.14 KGS/CM2)
  • Configuration: Vertical
  • Heat Zone Size: .58 Liters (35.3 cubic inches)
  • Heating Element: Metal (Tungsten)
  • Application: Multi-purpose




This laboratory vertical furnace is used for vacuum or inert gas treatment of metals and ceramics, reactive studies and small scale sintering studies.

Including:
  • Annealing
  • Brazing
  • Ceramic Firing
  • Co-Firing Ceramics
  • Compound Synthesis
  • Degassing
  • Densification
  • Electronic Ceramics
  • General Metals
  • General Thermo Processing
  • Hard Metal/Cermets
  • Heat Treating
  • Material Densification
  • Material Testing
  • Melting Point Determination
  • Powder Melting
  • Powder Processing
  • R & D Testing
  • Sintering



  • Ultra Clean Tungsten element and Tungsten/Molybdenum heat shields
  • Robust and Reliable Centorr Vacuum Industries' Design
  • Long Life from Rugged Centorr Vacuum Industries' Construction
  • Capable of many applications and processes
  • Highest Quality at an affordable price
  • Rapid heat up and cool down
  • Easy to operate
  • Simple to install (easy for customer to do)




Operating Temperature: 2000°C (3632°F) in vacuum, 2000°C (3632°F) in Argon, Nitrogen Gas, or Forming Gas.

Operating Pressure: 2 PSIG (0.14 KG/SQ. CM) positive pressure to 10-3 torr vacuum or 10-6 torr vacuum (dependent on model selection)

Ultimate Vacuum:
Standard Roughing Pump Model - 10-2 to 10-3 Torr (chamber clean, cold, empty and out gassed).
Diffusion Pump Model - 10-5 to 10-6 Torr (chamber clean, cold, empty and out gassed).
Turbo Molecular Pump Model - 10-5 to 10-6 Torr (chamber clean, cold, empty and out gassed).

Heat Up Time: 21°C to 2000°C, 176 minutes

Cool Down Time: 2000°C to 100°C, 19 minutes

Easy to Connect Water and Gas Connections



Standard Mechanical Pump Model: $44,400 USD

Diffusion Pump Model: $50,500 USD

Turbo Molecular Model: $56,400 USD




1 month from Purchase Order




Chamber: Double-walled, water-cooled 304L stainless steel chamber with water-cooled 304L stainless steel hinged lid.

Hot Zone: 3" (76.2 mm) dia. x 5" (127 mm) high tungsten mesh element with tungsten and molybdenum heat shields.

PUMPING SYSTEM ALTERNATIVES

Roughing Pump
Vacuum System:
Manual vacuum system with VRL 902111 vacuum instrument, Leybold Heraeus Model D8B, 198.2 LPM (7.0 CFM) direct drive mechanical pump and all necessary plumbing and manifolding.

Diffusion Pump
Vacuum System:
2" (50.8 mm) manual high vacuum system with Granville 330 vacuum instrument, Leybold Heraeus Model D8B (7.0 CFM) - 198.2 LPM) direct drive mechanical pump, Varian HS-2 diffusion pump, water baffle, high vacuum gate valve, and all necessary plumbing and manifolding.

Turbo Molecular
Vacuum Pumping System:
2" (50.8 mm) manual high vacuum system with Granville 330 vacuum instrument, high vacuum gate valve, Varian TMP V81, rated 77 L/S N2, Varian SH-100 dry scroll pump rated 90 L/M, and all necessary plumbing and manifolding.

Power Supply: 12 KVA SCR controlled with current limiting and soft start features-fully interlocked for failsafe operation.

Temperature Sensor: (2) Type "C" molybdenum sheathed thermocouples

Controller: Precision Digital PD-550. Programmable temperature controller with auto-tuning PID control. 1/16th DIN, two programs store 15 segments each for precision set-point control. Software for RS-485 serial communications included.

Overtemperature Controller
& Hot Zone Interlock:
Precision Digital PD-570 indicator controller. 1/16th DIN, hot zone temperature safety interlock. Software for RS-485 serial communications included.

Furnace Cabinet: Combination furnace, power supply, temperature control instrumentation, vacuum instrumentation, manually operated valving, enclosed in a painted steel cabinet.




Electrical: 12 KVA, 220V/1PH/50-60 Hz, 72 Amps Total. Connected load, 100 AMP Circuit

Water Supply: 2 GPM (7.5 LPM) at 70°F (21°C) max. 45 PSIG (3.1 Bar)
Max pH not to exceed 8.5.
Conductivity not to exceed 400 micro MHOS (Micro Siemens)
½" NPTM drain connection.

Water Return: 20 PSIG (1.4 Bar) Max. back pressure
½" NPTM drain connection

Inert Gas: Argon, nitrogen, or 4% H2/96% N2, approximately 0.2 cubic feet per backfill to atmospheric pressure, regulated at 10 PSIG.
¼" NPTF inlet connection
¼" NPTF exhaust connection

Pump Exhaust: ¾" (19.0 mm) ID Hose connection




Payment with Purchase Order
FOB: EX-WORKS (CVI Manufacturing Facility, Nashua, NH USA)
CVI Standard Terms and Conditions, Form TC-2000




Warranty, Manuals, Checkout and Field Service Warranty, Form C/VI-107




54" (137.1 CM) Long X 52" (132 CM) wide X 77"(195.5 CM) high
1250 Lbs. (567 kg)